Implementation of the qualities of radiodiagnostic: mammography

The objective of the present study was to evaluate the expanded uncertainty of the mammographic calibration process and present the result of the internal audit performed at the Laboratory of Radiological Sciences (LCR). The qualities of the mammographic beans that are references in the LCR, compris...

Full description

Saved in:
Bibliographic Details
Published inJournal of physics. Conference series Vol. 975; no. 1; pp. 12064 - 12068
Main Authors Pacífico, L C, Magalhães, L A G, Peixoto, J G P, Fernandes, E
Format Journal Article
LanguageEnglish
Published Bristol IOP Publishing 01.03.2018
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The objective of the present study was to evaluate the expanded uncertainty of the mammographic calibration process and present the result of the internal audit performed at the Laboratory of Radiological Sciences (LCR). The qualities of the mammographic beans that are references in the LCR, comprises two irradiation conditions: no-attenuated beam and attenuated beam. Both had satisfactory results, with an expanded uncertainty equals 2,1%. The internal audit was performed, and the degree of accordance with the ISO/IEC 17025 was evaluated. The result of the internal audit was satisfactory. We conclude that LCR can perform calibrations on mammography qualities for end users.