Implementation of the qualities of radiodiagnostic: mammography
The objective of the present study was to evaluate the expanded uncertainty of the mammographic calibration process and present the result of the internal audit performed at the Laboratory of Radiological Sciences (LCR). The qualities of the mammographic beans that are references in the LCR, compris...
Saved in:
Published in | Journal of physics. Conference series Vol. 975; no. 1; pp. 12064 - 12068 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
Bristol
IOP Publishing
01.03.2018
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | The objective of the present study was to evaluate the expanded uncertainty of the mammographic calibration process and present the result of the internal audit performed at the Laboratory of Radiological Sciences (LCR). The qualities of the mammographic beans that are references in the LCR, comprises two irradiation conditions: no-attenuated beam and attenuated beam. Both had satisfactory results, with an expanded uncertainty equals 2,1%. The internal audit was performed, and the degree of accordance with the ISO/IEC 17025 was evaluated. The result of the internal audit was satisfactory. We conclude that LCR can perform calibrations on mammography qualities for end users. |
---|---|