Repeatability of in-vitro optical quality measurements of intraocular lenses with a deflectometry technique effect of the toricity
To evaluate the repeatability of an optical device for measuring the Zernike coefficients of toric intraocular lenses (IOLs) and assess whether its toricity has any impact in its repeatability. An experienced technician used the NIMO TR1504 to measure the Zernike coefficients 30 times for an apertur...
Saved in:
Published in | International journal of ophthalmology Vol. 11; no. 7; pp. 1139 - 1144 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
China
International Journal of Ophthalmology Press
18.07.2018
Press of International Journal of Ophthalmology (IJO PRESS) |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | To evaluate the repeatability of an optical device for measuring the Zernike coefficients of toric intraocular lenses (IOLs) and assess whether its toricity has any impact in its repeatability.
An experienced technician used the NIMO TR1504 to measure the Zernike coefficients 30 times for an aperture of 4.50 mm for all lenses included. The IOLs included were divided into two group: toric and non-toric ones. The cylindrical powers of the toric lenses included in the present study were 1.00, 1.50, 2.25, 3.00 and 3.75 D. Finally, the repeatability of the NIMO TR1504 was described in terms of within subject standard deviation (Sw) and repeatability limit.
The Sw was smaller than 0.011 µm for both lens groups and all Zernike coefficients, and the difference between both groups was smaller than 0.004 µm for all Zernike coefficients. Regarding the repeatability limit, this value was smaller than 0.025 µm for the toric lens group, and smaller than 0.031 µm for the non-toric lens one for all Zernike coefficients. Furthermore, the maximum difference between both lens groups was 0.010 µm.
The repeatability of the NIMO TR1504 to measure the optical quality is high and independent of the lens toricity. These results reflect that this system is robust and could be used to measure the
optical quality of either toric or non-toric IOLs. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 2222-3959 2227-4898 |
DOI: | 10.18240/ijo.2018.07.11 |