The Use of TL Detectors in Dosimetry of Systemic Radiation Therapy

A method for determining absorbed doses to organs in systemic radiation therapy (SRT) is evaluated. The method, based on thermoluminescent (TL) dosimeters placed on the patient's skin, was validated and justified through a phantom study showing that the difference between measured (TL dosimeter...

Full description

Saved in:
Bibliographic Details
Published inActa oncologica Vol. 38; no. 2; pp. 189 - 196
Main Authors ASCHAN, A. C, TOIVONEN, M. J, LAMPINEN, J. S, TENHUNEN, M, KAIREMO, K. J. A, KORPPI-TOMMOLA, E. T, JEKUNEN, A. P, SIPILÄ, P, SAVOLAINEN, S. E
Format Journal Article
LanguageEnglish
Published Basingstoke Informa UK Ltd 1999
Taylor & Francis
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:A method for determining absorbed doses to organs in systemic radiation therapy (SRT) is evaluated. The method, based on thermoluminescent (TL) dosimeters placed on the patient's skin, was validated and justified through a phantom study showing that the difference between measured (TL dosimeters in the phantom) and derived (TL method) values is within 10%. Six radioimmunotherapy (RIT) patients with widespread intraperitoneal pseudomyxoma were also studied. In dose evaluations, special emphasis was on kidneys. In addition to the TL method, the absorbed doses to kidneys were calculated using MIRD formalism and a point dose kernel technique. We conclude that in SRT the described TL method can be used to estimate the absorbed doses to those critical organs near the body surface within 50% (1 SD).
ISSN:0284-186X
1651-226X
DOI:10.1080/028418699431609