Radiation Exposure in Vascular Angiographic Procedures

Abstract Purpose To evaluate dose reduction in vascular angiographic procedures by using fluoroscopy capture instead of digital subtraction angiography frames for documentation. Materials and Methods A total of 764 consecutive vascular interventional procedures performed over a period of 1 year were...

Full description

Saved in:
Bibliographic Details
Published inJournal of vascular and interventional radiology Vol. 23; no. 11; pp. 1487 - 1495
Main Authors Pitton, Michael Bernhard, MD, Kloeckner, Roman, MD, Schneider, Jens, MD, Ruckes, Christian, MS, Bersch, Anton, MD, Düber, Christoph, MD
Format Journal Article
LanguageEnglish
Published United States Elsevier Inc 01.11.2012
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Abstract Purpose To evaluate dose reduction in vascular angiographic procedures by using fluoroscopy capture instead of digital subtraction angiography frames for documentation. Materials and Methods A total of 764 consecutive vascular interventional procedures performed over a period of 1 year were retrospectively analyzed with respect to the fluoroscopy time and the resulting dose–area product (DAP), the DAP of the radiographic frames, and the overall DAP. Results A total of 70% of the total DAP was a result of the acquisition of radiographic frames, leaving only 30% being applied by fluoroscopy. Conclusions Fluoroscopy capture should be used for documentation whenever possible. A registry of radiation exposure should not only comprise a sufficiently large number of interventions but also different intervention types to allow the development of interventional reference levels.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1051-0443
1535-7732
DOI:10.1016/j.jvir.2012.05.048