Using strong injection solvents with 100% aqueous mobile phase in RP-LC
When samples are dissolved for HPLC analysis, organic solvents are often used to enhance the solubility of the sample components. However, when the diluent becomes significantly stronger than the mobile phase, peak shape distortions may develop during injection. This is a serious problem for weakly...
Saved in:
Published in | Journal of separation science Vol. 29; no. 18; pp. 2847 - 2852 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
Weinheim
WILEY-VCH Verlag
01.12.2006
WILEY‐VCH Verlag Wiley |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | When samples are dissolved for HPLC analysis, organic solvents are often used to enhance the solubility of the sample components. However, when the diluent becomes significantly stronger than the mobile phase, peak shape distortions may develop during injection. This is a serious problem for weakly retained analytes under 100% aqueous mobile phase conditions. Under these extreme conditions, even small amounts of solvent in the diluent can cause serious broadening or distortion effects. However, if the organic solvent used in the diluent elutes significantly after the analyte, the peak shape of the analyte will be relatively unaffected, even when a relatively strong solvent such as THF is used. This phenomenon is counterintuitive considering the usual practice of keeping the sample diluent as weak as possible. This report describes the potential analytical utility and limitations of this approach for the analysis of weakly retained analytes. |
---|---|
Bibliography: | ark:/67375/WNG-J9TMDX1J-D istex:98B2FD8AE0B77198C070ADB2CA5C64B3AF64E7C7 ArticleID:JSSC200600299 ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 1615-9306 1615-9314 |
DOI: | 10.1002/jssc.200600299 |