Simultaneous measurement of acoustic pressure and temperature at the focus of high intensity focused ultrasound in a tissue-mimicking phantom by using a fiber-optic hydrophone
The simultaneous measurement of the acoustic pressure and the temperature at the focus of high intensity focused ultrasound (HIFU) in a tissue-mimicking phantom were performed by using a dual sensing fiber-optic hydrophone (FOH). The HIFU pressure and the HIFU-induced temperature rise at the focus i...
Saved in:
Published in | Japanese Journal of Applied Physics Vol. 58; no. 9; pp. 90908 - 90912 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
Tokyo
IOP Publishing
01.09.2019
Japanese Journal of Applied Physics |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | The simultaneous measurement of the acoustic pressure and the temperature at the focus of high intensity focused ultrasound (HIFU) in a tissue-mimicking phantom were performed by using a dual sensing fiber-optic hydrophone (FOH). The HIFU pressure and the HIFU-induced temperature rise at the focus in the phantom were numerically simulated by using a MATLAB-based software package developed for HIFU simulation. The measurements of the HIFU-induced temperature rise showed good agreement with the simulations. It was also found that the presence of the FOH tip at the focus in the phantom did not affect the formation of a thermal lesion. |
---|---|
Bibliography: | JJAP-101527.R3 |
ISSN: | 0021-4922 1347-4065 |
DOI: | 10.7567/1347-4065/ab3881 |