Duplex pulsed Doppler echocardiography in mitral regurgitation
The accuracy of duplex pulsed Doppler echocardiography (d-PDE) for detecting mitral regurgitation was evaluated in 35 patients undergoing d-PDE, cardiac auscultation, and left ventriculography. With three transducer positions, the overall d-PDE sensitivity was 95%, specificity was 100%, positive pre...
Saved in:
Published in | Journal of clinical ultrasound Vol. 14; no. 2; p. 117 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
United States
01.02.1986
|
Subjects | |
Online Access | Get more information |
Cover
Loading…
Summary: | The accuracy of duplex pulsed Doppler echocardiography (d-PDE) for detecting mitral regurgitation was evaluated in 35 patients undergoing d-PDE, cardiac auscultation, and left ventriculography. With three transducer positions, the overall d-PDE sensitivity was 95%, specificity was 100%, positive predictive value was 100%, negative predictive value was 94%, and diagnostic accuracy was 97% compared with ventriculography. This technique was superior to auscultation (sensitivity 74%, specificity 94%, positive predictive value 93%, negative predictive value 75%, diagnostic accuracy 83%). No false-positive d-PDE results occurred, but discordant false-negative results occurred frequently among the three transducer positions. If discordant negative results are considered to be false negative, then d-PDE is both sensitive and specific when mitral regurgitation is defined as systolic spectral broadening in any one transducer position. |
---|---|
ISSN: | 0091-2751 |
DOI: | 10.1002/jcu.1870140207 |