Lowering radiation dose during dedicated colorectal cancer MDCT: comparison of low tube voltage and sinogram-affirmed iterative reconstruction at 80 kVp versus blended dual-energy images in a population of patients with low body mass index

Purpose To assess the diagnostic accuracy, cancer staging, image quality, and radiation dose of 80-kVp computed tomography (CT) images for patients with colorectal cancers (CRCs) using sinogram-affirmed iterative reconstruction (SAFIRE). Methods Sixty-four consecutive patients (mean weight 62.5  ±  ...

Full description

Saved in:
Bibliographic Details
Published inAbdominal imaging Vol. 40; no. 7; pp. 2867 - 2876
Main Authors Chen, Chiao-Yun, Hsu, Jui-Sheng, Jaw, Twei-Shiun, Kuo, Yu-Ting, Wu, Deng-Chyang, Lee, Chien-Hung, Shih, Ming-Chen Paul, Tsai, Tzu-Hsueh, Kuo, Chao-Hung, Chen, Yi-Ting, Yang, Li-Hwa, Liu, Gin-Chung
Format Journal Article
LanguageEnglish
Published New York Springer US 01.10.2015
Springer Nature B.V
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Purpose To assess the diagnostic accuracy, cancer staging, image quality, and radiation dose of 80-kVp computed tomography (CT) images for patients with colorectal cancers (CRCs) using sinogram-affirmed iterative reconstruction (SAFIRE). Methods Sixty-four consecutive patients (mean weight 62.5  ±  11.3 kg, mean BMI 24.1  ±  3.3 kg/m 2 ) with known CRC underwent dual-energy CT. Data were reconstructed as a weighted average (WA) 120-kVp dataset. Both filtered back projection (FBP) and SAFIRE were applied to reconstruct the WA 120-Kvp (Protocol A, B) and 80-kVp (Protocol C, D) image sets. The image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of the cancers, the normal reference tissues, and the effective dose for each protocol were assessed. The cancer detection, staging, and image quality were evaluated. Analysis of variance was used for statistical analysis. Results Compared with the FBP datasets at WA 120-kVp (Protocol A) and 80-kVp (Protocol C), the SAFIRE-reconstructed images (Protocols B, D) demonstrated significantly lower image noise ( P   <  0.0083). Protocol D yielded significantly higher CNRs and SNRs for the CRCs and normal reference tissues than did Protocols A and C ( P   <  0.0083). Protocol D also exhibited a significantly higher CNR for the CRC and some normal reference tissues than did Protocol B ( P   <  0.0083). For hypovascular liver metastases ( n   =  10), Protocol D yielded better SNRs and significantly higher CNRs than did Protocol A ( P   <  0.0083). Overall, accuracy for tumor staging and liver metastasis was 95.3% (61/64) and 100%, respectively, in all of the 4 protocols. The mean effective dose decreased 41% from the WA 120-kVp to the 80-kVp protocols (6.23 vs. 3.68 mSv). Conclusions The 80-kVp technique with SAFIRE provided high SNR, high CNR, and good accuracy for staging in nonobese patients with CRC. Our study results should be extrapolated to patient populations with a high BMI with caution. Further studies of high BMI patients are therefore warranted.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0942-8925
2366-004X
1432-0509
2366-0058
DOI:10.1007/s00261-015-0412-1