The PediSedate® device, a novel approach to pediatric sedation that provides distraction and inhaled nitrous oxide: clinical evaluation in a large case series
Summary Background: Pediatric sedation is of paramount importance but can be challenging. Fear and anticipatory anxiety before invasive procedures often lead to uncooperativeness. A novel device (PediSedate®) provides sedation through a combination of inhaled nitrous oxide and distraction (video ga...
Saved in:
Published in | Pediatric anesthesia Vol. 17; no. 2; pp. 162 - 166 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
Oxford, UK
Blackwell Publishing Ltd
01.02.2007
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Summary
Background: Pediatric sedation is of paramount importance but can be challenging. Fear and anticipatory anxiety before invasive procedures often lead to uncooperativeness. A novel device (PediSedate®) provides sedation through a combination of inhaled nitrous oxide and distraction (video game). We evaluated the acceptability and safety of the PediSedate® device in children.
Methods: We enrolled children between 3 and 9 years old who were scheduled to undergo surgical procedures that required general inhalational anesthesia. After the device was applied, he/she played a video game while listening to the audio portion of the game through the earphones. Nitrous oxide in oxygen was administered via the nasal piece of the headset starting at 50% and increasing to 70%, in 10% increments every 8 min. Treatment failures, vital signs, arterial oxygen saturation, depth of sedation, airway patency, side effects, acceptance of the device and parental satisfaction were all evaluated.
Results: Of 100 children included, treatment failure occurred in 18% mainly because of poor tolerance of the device. At least 96% of the children who completed the study exhibited an excellent degree of sedation, 22% had side effects, and none experienced serious airway obstruction. Nausea and vomiting were the most common side effects and no patients had hemodynamic instability.
Conclusions: The PediSedate® device combines nonpharmacologic with pharmacologic methods of sedation. Most of the children we evaluated were able to tolerate the PediSedate® device and achieved an adequate degree of sedation. |
---|---|
Bibliography: | ark:/67375/WNG-S0GZ20NJ-V ArticleID:PAN2091 istex:29300FAF27077F5AB5CC094C79DF187250E032F2 ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 |
ISSN: | 1155-5645 1460-9592 |
DOI: | 10.1111/j.1460-9592.2006.02091.x |