Optical Biometry and IOL Calculation in a Commercially Available Optical Coherence Tomography Device and Comparison With Pentacam AXL
Optical devices are the gold standard for ocular biometry; however, they are unable to obtain high-quality optical coherence tomography (OCT) images. The current study aimed to evaluate ocular measurements and intraocular lens (IOL) calculation used in an anterior/posterior segment OCT device and to...
Saved in:
Published in | American journal of ophthalmology Vol. 246; pp. 236 - 241 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
United States
Elsevier Inc
01.02.2023
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Optical devices are the gold standard for ocular biometry; however, they are unable to obtain high-quality optical coherence tomography (OCT) images. The current study aimed to evaluate ocular measurements and intraocular lens (IOL) calculation used in an anterior/posterior segment OCT device and to compare the results with those of a validated biometer.
Prospective evaluation of a diagnostic tool.
This study enrolled healthy subjects at the Hygeia Clinic, Gdańsk, Poland, between October 2021 and November 2021. All individuals had ocular biometry measured with a validated biometer (Pentacam AXL) and with a new module of an anterior/posterior segment OCT device (Revo 80, Optopol Technologies). All IOL calculations were performed for the right eye with keratometric values from the Pentacam for one IOL: the Alcon AcrySof IQ SN60WF, with plano target setting.
The mean age of the 144 participants was 25.23 ± 7.15 years. The axial length measured with Revo was longer than with Pentacam AXL (24.08 ± 1.13 vs 23.98 ± 1.13; P < .0001), a 0.10 ± 0.04 mm difference. This translated into a significantly lower IOL power to achieve emmetropia for all formulas (–0.34 ± 0.15, –0.32 ± 0.13, –0.34 ± 0.19, and –0.30 ± 0.15 for the Hoffer Q, Holladay I, Haigis, and SRK/T formulas, respectively). The study showed high agreement between the devices: nearly 90% of eyes were within ±0.50 diopters for all of the analyzed formulas (r > 0.99).
The present study demonstrates that the results of IOL calculation with the OCT biometer have a very strong correlation with those obtained with the Pentacam AXL; however, axial length measurements and calculated IOL power cannot be considered interchangeable. |
---|---|