A high-precision angular control system for HIFU calibration
► We designed a high precision angular alignment system with a proper protocol for ultrasound field measurement. ► We examined the performance of the proposed design and compared it with that of commercial equipment. ► The obtained accuracy of alignment error with the proposed system was less than 0...
Saved in:
Published in | Ultrasonics Vol. 53; no. 1; pp. 45 - 52 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | English |
Published |
Netherlands
Elsevier B.V
01.01.2013
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | ► We designed a high precision angular alignment system with a proper protocol for ultrasound field measurement. ► We examined the performance of the proposed design and compared it with that of commercial equipment. ► The obtained accuracy of alignment error with the proposed system was less than 0.01radian.
A design of high-precision angular position control system for calibrating high intensity focused ultrasound (HIFU) is presented with alignment procedures. Two independent angular controls are achieved by combining a worm gear and a belt gear system. The proposed system verifies alignment by comparing simulation data and experimental data with three different transducers and two different types of hydrophones. The performance of the proposed system is compared to that of a commercial system. The results indicate that the proposed system provides high precision angular alignment (e.g., <0.01radians) with robust reproducibility regardless of the hydrophone type. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0041-624X 1874-9968 |
DOI: | 10.1016/j.ultras.2012.03.012 |