Hemodynamic evaluation of the heart with a nonfluoroscopic electromechanical mapping technique

Clinical cardiac volumetric measurement techniques are essential for assessing cardiac performance but produce significant inaccuracies in extrapolation of the volume of a three-dimensional (3D) object from two-dimensional images and lack the ability to associate cardiac electrical and mechanical ac...

Full description

Saved in:
Bibliographic Details
Published inCirculation (New York, N.Y.) Vol. 96; no. 10; pp. 3672 - 3680
Main Authors GEPSTEIN, L, HAYAM, G, SHPUN, S, BEN-HAIM, S. A
Format Journal Article
LanguageEnglish
Published Hagerstown, MD Lippincott Williams & Wilkins 18.11.1997
American Heart Association, Inc
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Clinical cardiac volumetric measurement techniques are essential for assessing cardiac performance but produce significant inaccuracies in extrapolation of the volume of a three-dimensional (3D) object from two-dimensional images and lack the ability to associate cardiac electrical and mechanical activities. In this study, we tested the accuracy of cardiac volumetric measurements using a new catheter-based system. The system uses magnetic technology to accurately locate a special catheter at a frequency of 125 Hz and is currently used in the field of electrophysiology, in which activation maps are superimposed on the 3D geometry of the cardiac chamber. The mapping procedure is based on sequentially acquiring the location of the tip and local electrogram while in contact with the endocardium. The 3D geometry of the chamber is reconstructed in real time, and its volume could be calculated at every time step (8 ms). The volumetric measurements of the system were found to be highly accurate for simple phantoms (mean+/-SEM deviation, 2.3+/-1.1%), left ventricular casts (9.6+/-1.3%), and a dynamic test jig. In addition, left ventricular volumes of 12 swine were measured. Intraobserver and interobserver variabilities were found to be minimal (ejection fraction, 6.5+/-1.9% and 7.1+/-2.0%; stroke volume, 4.5+/-1.0% and 11.3+/-2.4%). Comparison with the thermodilution method for measuring stroke volume showed an average deviation of 8.1+/-2.2%. Typical pressure-volume loops were also obtained. The new mapping image provides, for the first time, simultaneous information regarding cardiac mechanics, hemodynamics, and electrical properties. Furthermore, all this information is achieved without the use of fluoroscopy, contrast medium, or complicated image processing.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0009-7322
1524-4539
DOI:10.1161/01.CIR.96.10.3672