The determination of measurement errors in impedance spectroscopy

The paper deals with the methodology, quantification and comparison of measurement deviations for two types of the ceramic shard. The ceramic material has better polarization properties than concrete material. For impedance measurement the ceramic shard is more appropriate. The IS method diagnostics...

Full description

Saved in:
Bibliographic Details
Published inIOP conference series. Materials Science and Engineering Vol. 583; no. 1; pp. 12030 - 12034
Main Authors Lunak, M, Kusak, I, Chobola, Z
Format Journal Article
LanguageEnglish
Published Bristol IOP Publishing 01.07.2019
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The paper deals with the methodology, quantification and comparison of measurement deviations for two types of the ceramic shard. The ceramic material has better polarization properties than concrete material. For impedance measurement the ceramic shard is more appropriate. The IS method diagnostics based on processing of measured data which aims evaluating the absolute error and the relative error (deviation) of the loss factor values. Results show the spectrum parts which are useful in the lower frequency range only. When Agilent 33220A and Agilent 54645A are used to carry out the IS, the most important factor from the viewpoint of the tgδ(f) determination accuracy is the accuracy of the phase shift between the excitation signal and that at the measured system output. Other factors are the measurement accuracy of the electric capacity and resistance of the oscilloscope input channel.
ISSN:1757-8981
1757-899X
DOI:10.1088/1757-899X/583/1/012030