Development of an Automated Diffusion Scrubber-Conductometry System for Measuring Atmospheric Ammonia
A semi-continuous and automated method for quantifying atmospheric ammonia at the parts per billion level has been developed. The instrument consists of a high efficiency diffusion scrubber, an electrolytic on-line anion exchange device, and a conductivity detector. Water soluble gases in sampled ai...
Saved in:
Published in | Bulletin of the Korean Chemical Society Vol. 32; no. 6; pp. 2039 - 2044 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
대한화학회
20.06.2011
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | A semi-continuous and automated method for quantifying atmospheric ammonia at the parts per billion level has been developed. The instrument consists of a high efficiency diffusion scrubber, an electrolytic on-line anion exchange device, and a conductivity detector. Water soluble gases in sampled air diffuse through the porous membrane and are absorbed in an absorbing solution. Interferences are eliminated by using an anion exchange devises. The electrical conductivity of the solution is measured without chromatographic separation.
The collection efficiency was over 99%. Over the 0-200 ppbv concentration range, the calibration was linear with r2 = 0.99. The lower limit of detection was 0.09 ppbv. A parallel analysis of Seoul air over several days using this method and a diffusion scrubber coupled to an ion chromatography system showed acceptable agreement, r2 = 0.940 (n = 686). This method can be applied for ambient air monitoring of ammonia. KCI Citation Count: 1 |
---|---|
Bibliography: | G704-000067.2011.32.6.023 |
ISSN: | 0253-2964 1229-5949 |