Final 5-Year Results in Unselected Patients Implanted With a Thin-Strut, Platinum-Chromium, Everolimus-Eluting Stent (from the PROMUS Element Plus US Post-Approval Study)

PROMUS Element (PE) Plus post-approval study was a large prospective, observational, all-comers study designed to evaluate the safety and performance of a thin-strut platinum chromium everolimus-eluting PE Plus stent in everyday clinical practice. A total of 2,683 “real-world” patients with limited...

Full description

Saved in:
Bibliographic Details
Published inThe American journal of cardiology Vol. 123; no. 11; pp. 1765 - 1771
Main Authors Kandzari, David E., Amjadi, Nima, Caputo, Christopher, Rowe, Steven K., Chen, Henry, Williams, Jerome, Tamboli, Hoshedar P., Underwood, Paul L., Allocco, Dominic J., Meredith, Ian T.
Format Journal Article
LanguageEnglish
Published United States Elsevier Inc 01.06.2019
Elsevier Limited
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:PROMUS Element (PE) Plus post-approval study was a large prospective, observational, all-comers study designed to evaluate the safety and performance of a thin-strut platinum chromium everolimus-eluting PE Plus stent in everyday clinical practice. A total of 2,683 “real-world” patients with limited clinical or anatomic exclusion criteria were enrolled at 52 clinical sites in the United States. The study met its primary end point of 12-month cardiac death or myocardial infarction (CD/MI) compared with a prespecified performance goal (p <0.0001). Five-year clinical outcomes were evaluated in overall PE Plus post-approval study patients and high-risk subgroups. During the 5-year follow-up period, CD/MI and stent thrombosis related to the PE Plus stent occurred in 9% and 2.2% overall patients, respectively. The reported all-cause mortality rate was 15%, with 7% classified as cardiac-related. A total of 18% patients underwent target vessel revascularization, and 11% were reported as target lesion revascularization. The rates of PE Plus stent-related CD/MI remained low in patients with medically treated diabetes (13%), small vessels (9%), and long stents (10%). PE Plus stent-related thrombosis through 5 years in patients with diabetes, small vessels and long stents was 4.2%, 2.2%, and 2.6%, respectively. The occurrence of target lesion revascularization was numerically higher in patients with diabetes (16%) and long lesions (18%) than the small vessels subgroup (11%). In conclusion, the final 5-year results establish the long-term safety and efficacy of the PE Plus stent in a broad, unselected patient population representative of “real-world” clinical practice.