Reliability of continuous jugular venous bulb hemoglobin oxygen saturation during cardiac surgery
Objective: To evaluate the accuracy and reliability of continuous measurement of jugular venous bulb hemoglobin oxygen saturation (SjvO 2) with a fiberoptic catheter (SjvO 2OX) during cardiac surgery versus simultaneous paired measurements of hemoglobin oxygen saturation by the Hemoximeter (SjvO 2HE...
Saved in:
Published in | Journal of cardiothoracic and vascular anesthesia Vol. 13; no. 3; pp. 276 - 279 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
Philadelphia, PA
Elsevier Inc
01.06.1999
Elsevier |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Objective:
To evaluate the accuracy and reliability of continuous measurement of jugular venous bulb hemoglobin oxygen saturation (SjvO
2) with a fiberoptic catheter (SjvO
2OX) during cardiac surgery versus simultaneous paired measurements of hemoglobin oxygen saturation by the Hemoximeter (SjvO
2HEM; Radiometer, Copenhagen, Denmark) and indirect estimations of hemoglobin oxygen saturation from measurements of partial pressure of oxygen in blood gases (SjvO
2BG).
Design:
A prospective study.
Setting:
American Hellenic Educational Progressive Association General Hospital, University Hospital of Thessaloniki, Greece.
Patients:
Thirty patients undergoing elective aortocoronary artery bypass surgery.
Interventions:
In addition to routine pressure monitoring, a 4F fiberoptic catheter was placed in the left jugular bulb by a retrograde internal jugular vein approach and SjvO
2OX was continuously measured. Before insertion, each catheter was calibrated in vitro.
Measurements and Main Results:
One hundred twelve simultaneous paired recordings between SjvO
2OX and SjvO
2BG were performed to define the accuracy of SjvO
2OX to SjvO
2BG. Sixty-one of 112 simultaneous paired recordings between SjvO
2OX and SjvO
2HEM and SjVO
2HEM and SjvO
2BG were performed to define the accuracy of SjvO
2OX to the reference SjVO
2HEM and the reliability of the SjVO
2BG measurement to SjVO
2HEM. The fiberoptic catheter readings varied from underestimating to overestimating hemoglobin saturation by a mean of −5.35% to +9.67% and of −3.22% to +7.81% versus Blood Gas Analyzer (Ciba-Corning) and Co-Oximeter (OSM 2b Hemoximeter, Radiometer) values, respectively. The mean underestimation and overestimation of CoOximeter versus Blood Gas Analyzer values were −3.18% and +4.17%, respectively.
Conclusion:
SjvO
2 values obtained continuously from a jugular venous bulb fiberoptic catheter may give relatively accurate readings provided they are duly interpreted and errors caused by wall artifact or blood sampling are avoided. |
---|---|