A miniaturized carbon dioxide gas sensor based on sensing of pH-sensitive hydrogel swelling with a pressure sensor
A measurement concept has been realized for the detection of carbon dioxide, where the CO(2) induced pressure generation by an enclosed pH-sensitive hydrogel is measured with a micro pressure sensor. The application of the sensor is the quantification of the partial pressure of CO(2) (Pco(2)) in the...
Saved in:
Published in | Biomedical microdevices Vol. 7; no. 3; pp. 197 - 204 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
United States
Springer Nature B.V
01.09.2005
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | A measurement concept has been realized for the detection of carbon dioxide, where the CO(2) induced pressure generation by an enclosed pH-sensitive hydrogel is measured with a micro pressure sensor. The application of the sensor is the quantification of the partial pressure of CO(2) (Pco(2)) in the stomach as diagnosis for gastrointestinal ischemia. The principle is put to the proof by examining the sensor response to changes in Pco(2). Furthermore, the response time, temperature-sensitivity and resolution are determined. The sensor responds well to changes in Pco(2) with a maximum pressure generation of 0.29 x 10(5) Pa at 20 kPa CO(2). The 90% response time varies between 1.5 and 4.5 minutes at 37( composite function)C. The sensor shows a linear temperature-sensitivity which can easily be compensated for, and enables detection of Pco(2) changes as small as 0.5 kPa CO(2). |
---|---|
Bibliography: | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 ObjectType-Article-1 ObjectType-Feature-2 |
ISSN: | 1387-2176 1572-8781 |
DOI: | 10.1007/s10544-005-3026-5 |