Endless loop tachycardia started by an atrial premature complex in a patient with a dual chamber pacemaker
In a patient with a dual chamber pacemaker that senses in both the atrium and ventricle (VDD, DDD), a ventricular depolarization temporally displaced from a P wave can cause retrograde atrial activation and initiate an endless loop pacemaker-mediated tachycardia. A case in which an endless loop tach...
Saved in:
Published in | Journal of the American College of Cardiology Vol. 5; no. 3; pp. 707 - 710 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
New York, NY
Elsevier Inc
01.03.1985
Elsevier Science |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | In a patient with a dual chamber pacemaker that senses in both the atrium and ventricle (VDD, DDD), a ventricular depolarization temporally displaced from a P wave can cause retrograde atrial activation and initiate an endless loop pacemaker-mediated tachycardia. A case in which an endless loop tachycardia was initiated by an end-diastolic atrial premature complex is reviewed. Retrograde conduction occurred because of the change in the temporal relation of atrial sensing and atrioventricular (AV) node depolarization. The implanted pacemaker did not have the capability of atrial refractory programmability. Atrial refractory interval extension, which occurs in this model after a ventricular premature complex to protect against a retrograde P wave, was not invoked since the tachycardia was begun by an atrial rather than a ventricular premature complex. The tachycardia was controlled by shortening the programmable AV delay. The mechanism of tachycardia induction and its management are outlined. Atrial refractory programmability is required in all VDD or DDD pacemakers. |
---|---|
Bibliography: | ObjectType-Case Study-2 SourceType-Scholarly Journals-1 ObjectType-Feature-4 content type line 23 ObjectType-Report-1 ObjectType-Article-3 |
ISSN: | 0735-1097 1558-3597 |