Does cone-beam CT convex triangular field of view influence the image shape distortion of high-density materials?

To assess the distortion of high-density materials using two CBCT devices presenting convex triangular and cylindrical fields of view (FOVs). Four high-density cylinders were individually placed in a polymethylmethacrylate phantom. 192 CBCT scans were acquired using the convex triangular and cylindr...

Full description

Saved in:
Bibliographic Details
Published inDento-maxillo-facial radiology Vol. 52; no. 8; p. 20230029
Main Authors Cascante-Sequeira, Deivi, Coelho-Silva, Fernanda, Lopes Rosado, Lucas P, Lucca, Luiza Valdemarca, Queiroz Freitas, Deborah, Lins de-Azevedo-Vaz, Sergio, Haiter-Neto, Francisco
Format Journal Article
LanguageEnglish
Published England The British Institute of Radiology 01.11.2023
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:To assess the distortion of high-density materials using two CBCT devices presenting convex triangular and cylindrical fields of view (FOVs). Four high-density cylinders were individually placed in a polymethylmethacrylate phantom. 192 CBCT scans were acquired using the convex triangular and cylindrical FOVs of Veraviewepocs R100 (R100) and Veraview X800 (X800) devices. Using Horos 's software, two oral radiologists determined the cylinders' horizontal and vertical dimensional alterations. Nine oral radiologists subjectively identified each cylinder's axial shape distortion. Statistical analysis comprised Multiway ANOVA (α = 5%), and the Kruskal-Wallis test. The distortion in the axial plane was greater in the convex triangular FOVs for both devices in almost all the materials ( < 0.05). The evaluators subjectively identified a shape distortion in both FOVs for R100 device ( < 0.001), while no distortion was identified for X800 device ( = 0.620). A vertical magnification of all materials was observed in both FOVs for both devices ( < 0.05). No differences among vertical regions ( = 0.988) nor FOVs ( = 0.544) were found for the R100 device, while all materials showed higher magnification in all regions in the cylindrical FOV ( < 0.001) of the X800 device. The convex triangular FOV influenced the axial distortion of the high-density materials in both devices. A vertical magnification was observed in both FOVs of both devices, but it was greater in the cylindrical FOV of the X800 device.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0250-832X
1476-542X
DOI:10.1259/dmfr.20230029