The performance of commercial photodiodes for dosimetry in mammography
This paper reports the results of an investigation carried out to determine the suitability of using a commercially available photodiode for dosimetry in mammography thereby providing a low cost dosemeter with a direct reading of the radiation dose. A mammographic X-ray generator (Siemens Mammomat 1...
Saved in:
Published in | Radiation protection dosimetry Vol. 115; no. 1-4; p. 391 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
England
01.01.2005
|
Subjects | |
Online Access | Get more information |
Cover
Loading…
Summary: | This paper reports the results of an investigation carried out to determine the suitability of using a commercially available photodiode for dosimetry in mammography thereby providing a low cost dosemeter with a direct reading of the radiation dose. A mammographic X-ray generator (Siemens Mammomat 1000) with tube potential in the 23-30 kV range and a constant potential X-ray generator (Pantak) with a W/Mo anode/filter combination were used in this study. The results showed that the photodiode response is highly linear within mammographic dose ranges and that the energy dependence was <3% at tube potentials of 25-30 kV. Good agreement was observed between the incident air kerma measured with both the photodiode and the ionization chamber. These results show the viability of using the photodiode as a dosemeter system in mammography. |
---|---|
ISSN: | 0144-8420 |
DOI: | 10.1093/rpd/nci045 |