Cerebral time domain-NIRS: reproducibility analysis, optical properties, hemoglobin species and tissue oxygen saturation in a cohort of adult subjects
The reproducibility of cerebral time-domain near-infrared spectroscopy (TD-NIRS) has not been investigated so far. Besides, reference intervals of cerebral optical properties, of absolute concentrations of deoxygenated-hemoglobin (HbR), oxygenated-hemoglobin (HbO), total hemoglobin (HbT) and tissue...
Saved in:
Published in | Biomedical optics express Vol. 8; no. 11; pp. 4987 - 5000 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | English |
Published |
United States
Optical Society of America
01.11.2017
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | The reproducibility of cerebral time-domain near-infrared spectroscopy (TD-NIRS) has not been investigated so far. Besides, reference intervals of cerebral optical properties, of absolute concentrations of deoxygenated-hemoglobin (HbR), oxygenated-hemoglobin (HbO), total hemoglobin (HbT) and tissue oxygen saturation (StO
) and their variability have not been reported. We have addressed these issues on a sample of 88 adult healthy subjects. TD-NIRS measurements at 690, 785, 830 nm were fitted with the diffusion model for semi-infinite homogenous media. Reproducibility, performed on 3 measurements at 5 minutes intervals, ranges from 1.8 to 6.9% for each of the hemoglobin species. The mean ± SD global values of HbR, HbO, HbT, StO
are respectively 24 ± 7 μM, 33.3 ± 9.5 μM, 57.4 ± 15.8 μM, 58 ± 4.2%. StO
displays the narrowest range of variability across brain regions. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 2156-7085 2156-7085 |
DOI: | 10.1364/BOE.8.004987 |