Analysis of retrieved STRYDE nails

Aims The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices. Methods We collected ten nails removed from six patients, t...

Full description

Saved in:
Bibliographic Details
Published inBone & joint open Vol. 2; no. 8; pp. 599 - 610
Main Authors Hothi, Harry, Bergiers, Sean, Henckel, Johann, Iliadis, Alexis D., Goodier, William David, Wright, Jonathan, Skinner, John, Calder, Peter, Hart, Alister J.
Format Journal Article
LanguageEnglish
Published London The British Editorial Society of Bone & Joint Surgery 01.08.2021
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Aims The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices. Methods We collected ten nails removed from six patients, together with patient clinical data and plain radiograph imaging. We performed macro- and microscopic analysis of all surfaces and graded the presence of corrosion using validated semiquantitative scoring methods. We determined the elemental composition of surface debris using energy dispersive x-ray spectroscopy (EDS) and used metrology analysis to characterize the surface adjacent to the extendable junctions. Results All nails were removed at the end of treatment, having achieved their intended lengthening (20 mm to 65 mm) and after regenerate consolidation. All nails had evidence of corrosion localized to the screw holes and the extendable junctions; corrosion was graded as moderate at the junction of one nail and severe at the junctions of five nails. EDS analysis showed surface deposits to be chromium rich. Plain radiographs showed cortical thickening and osteolysis around the junction of six nails, corresponding to the same nails with moderate – severe junction corrosion. Conclusion We found, in fully united bones, evidence of cortical thickening and osteolysis that appeared to be associated with corrosion at the extendable junction; when corrosion was present, cortical thickening was adjacent to this junction. Further work, with greater numbers of retrievals, is required to fully understand this association between corrosion and bony changes, and the influencing surgeon, implant, and patient factors involved. Cite this article: Bone Jt Open 2021;2(8):599–610.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:2633-1462
2633-1462
DOI:10.1302/2633-1462.28.BJO-2021-0126