An investigation of inter-shock timing and electrode placement for double-sequential defibrillation

Double-Sequential Defibrillation (DSD) is the near-simultaneous use of two defibrillators to treat refractory VF. We hypothesized that (1) risk of DSD-associated defibrillator damage depends on shock vector and (2) the efficacy of DSD depends on inter-shock time. Part 1: risk of defibrillator damage...

Full description

Saved in:
Bibliographic Details
Published inResuscitation Vol. 140; pp. 194 - 200
Main Authors Taylor, Tyson G., Melnick, Sharon B., Chapman, Fred W., Walcott, Gregory P.
Format Journal Article
LanguageEnglish
Published Ireland Elsevier B.V 01.07.2019
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Double-Sequential Defibrillation (DSD) is the near-simultaneous use of two defibrillators to treat refractory VF. We hypothesized that (1) risk of DSD-associated defibrillator damage depends on shock vector and (2) the efficacy of DSD depends on inter-shock time. Part 1: risk of defibrillator damage was assessed in three anaesthetized pigs by applying two sets of defibrillation electrodes in six different configurations (near-orthogonal or near-parallel vectors). Ten 360J shocks were delivered from one set of pads and peak voltage was measured across the second set. Part 2: the dependence of DSD efficacy on inter-shock time was assessed in ten anaesthetized pigs. Electrodes were applied in lateral-lateral (LL) and anterior-posterior positions. Control (LL Stacked Shocks; one vector, two shocks ∼10 s apart) and DSD therapies (Overlapping, 10 ms, 50 ms, 100 ms, 200 ms, 500 ms, 1000 ms apart) were tested in a block randomized design treating electrically-induced VF (n = ∼89 VF episodes/therapy). Shock energies were selected to achieve 25% shock success for a single LL shock. Part 1: peak voltage delivered was 1833 ± 48 V (mean ± 95%CI). Peak voltage exposure was, on average, 10-fold higher for parallel than orthogonal vectors (p < 0.0001). Part 2: DSD efficacy compared to Stacked LL shocks was higher for Overlapping, 10 ms, and 100 ms (p < 0.05); lower at 50 ms (p < 0.05); and not different at 200 ms or longer inter-shock times. Risk of DSD-associated defibrillator damage can be mitigated by using near-orthogonal shock vectors. DSD efficacy is highly dependent on the inter-shock time and can be better, worse, or no different than stacked shocks from a single vector. University of Alabama at Birmingham Institutional Animal Care and Use Committee (IACUC) Protocol Number 06860.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0300-9572
1873-1570
DOI:10.1016/j.resuscitation.2019.04.042