Validation and inter-laboratory study of selective hydride generation for fast screening of inorganic arsenic in seafood

It is advisable to monitor and regulate inorganic arsenic (iAs) in food and feedstuff. This work describes an update and validation of a method of selective hydride generation (HG) with inductively coupled plasma mass spectrometry (ICP-MS) for high-throughput screening of iAs content in seafood samp...

Full description

Saved in:
Bibliographic Details
Published inAnalytica chimica acta Vol. 1049; pp. 20 - 28
Main Authors Marschner, Karel, Pétursdóttir, Ásta H., Bücker, Patrick, Raab, Andrea, Feldmann, Jörg, Mester, Zoltán, Matoušek, Tomáš, Musil, Stanislav
Format Journal Article
LanguageEnglish
Published Netherlands Elsevier B.V 21.02.2019
Elsevier BV
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:It is advisable to monitor and regulate inorganic arsenic (iAs) in food and feedstuff. This work describes an update and validation of a method of selective hydride generation (HG) with inductively coupled plasma mass spectrometry (ICP-MS) for high-throughput screening of iAs content in seafood samples after microwave-assisted extraction with diluted nitric acid and hydrogen peroxide. High concentration of HCl (8 M) for HG along with hydrogen peroxide in samples of a same concentration as used for extraction leads to a selective conversion of iAs to volatile arsine that is released and transported to the detector. A minor contribution from methylarsonate (≈20% to iAs) was found, while HG from dimethylarsinate, trimethylarsine oxide is substantially suppressed (less than 1% to iAs). Methodology was applied to Certified Reference Materials (CRMs) TORT-3, DORM-3, DORM-4, DOLT-4, DOLT-5, PRON-1, SQID-1 and ERM-CE278k, in some of them iAs has been determined for the first time, and to various seaweed samples from a local store. The results were always compared with a reference method and selectivity of iAs determination was evaluated. An inter-laboratory reproducibility was tested by comparative analyses of six fish and four seaweed samples in three European laboratories, with good agreement of the results. The method of HG-ICP-MS is sensitive (limit of detection 2 μg kg−1 iAs), well suited for screening of large number of samples and selective at iAs concentration levels at which maximum limits are expected to be set into EU legislation for marine samples. [Display omitted] •Selective hydride generation (no chromatography) improved for iAs determination.•The method was validated for seafood samples with complex As speciation.•Sufficient selectivity and good inter-laboratory reproducibility was achieved.•The method is suited for fast screening of large sample sets in control laboratories.•New data on iAs content in Certified Reference Materials presented.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0003-2670
1873-4324
DOI:10.1016/j.aca.2018.11.036