Comparison of Corneal Tomography: Repeatability, Precision, Misalignment, Mean Elevation, and Mean Pachymetry
To compare corneal tomography and its statistical uncertainty for measurements obtained by three clinically used corneal tomographers: A Scheimpflug camera (Pentacam HR), a swept source optical coherence tomography system (CASIA SS-1000), and Placido ring imaging (TMS-5). Repeated measurements with...
Saved in:
Published in | Current eye research Vol. 43; no. 6; pp. 709 - 716 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
England
03.06.2018
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | To compare corneal tomography and its statistical uncertainty for measurements obtained by three clinically used corneal tomographers: A Scheimpflug camera (Pentacam HR), a swept source optical coherence tomography system (CASIA SS-1000), and Placido ring imaging (TMS-5).
Repeated measurements with all three devices on 34 normal eyes were used to estimate the repeatability, precision, and mean values of corneal elevation and pachymetry within 8 mm diameter. The repeatability (standard deviation) was calculated for each data point of the corneal elevation data-maps of anterior and posterior cornea as well as for the pachymetry data-maps. Uncertainty on the position of the eye at each measurement might contribute to the differences between elevation data-maps. To take this into account, we defined the precision as the standard deviation for the elevation data-maps of anterior and posterior cornea after correction of misalignment-effects (rotation, translation). The mean elevation and pachymetry data-maps were fitted with Zernike polynomials for interdevice-comparison.
Pentacam HR offered the best repeatability and precision for the anterior corneal elevation (<3 and <1.6 μm, respectively). CASIA SS-1000 offered good repeatability and precision with high resolution for posterior corneal elevation, and the best repeatability for pachymetry (<3 μm). TMS-5 measured anterior elevation with similar repeatability to CASIA SS-1000 (<6 μm). The data-maps of the three tomographers could not be used interchangeably. The largest differences were observed for pachymetry and posterior corneal elevation data-maps.
Misalignment limited the repeatability of TMS-5 and Pentacam HR, but had little influence on the repeatability of CASIA SS-1000. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0271-3683 1460-2202 |
DOI: | 10.1080/02713683.2018.1441873 |