Purpose To evaluate the coagulative performance of a 21-gauge (G) internally cooled radiofrequency (RF) electrode using ex vivo and in vivo rat liver. Materials and Methods We developed a prototype of 21-G internally cooled monopolar RF electrode with 5.0 mm active tip length. The ablative zone size...

Full description

Saved in:
Bibliographic Details
Published inCardiovascular and interventional radiology Vol. 44; no. 1; pp. 150 - 155
Main Authors Takaki, Haruyuki, Kodama, Yuki, Miyamoto, Hisao, Iijima, Yuto, Kikuchi, Yoshinari, Kodama, Hiroshi, Ueshima, Eisuke, Nakasho, Keiji, Wada, Reona, Yamakado, Koichiro
Format Journal Article
LanguageEnglish
Published New York Springer US 2021
Springer Nature B.V
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Purpose To evaluate the coagulative performance of a 21-gauge (G) internally cooled radiofrequency (RF) electrode using ex vivo and in vivo rat liver. Materials and Methods We developed a prototype of 21-G internally cooled monopolar RF electrode with 5.0 mm active tip length. The ablative zone size created by this electrode was evaluated in ex vivo and in vivo rat liver. Five RF powers (3 W, 5 W, 7 W, 9 W, and 11 W) were applied with and without circulation of chilled water within the electrode. The ablation zone sizes were compared. Histopathological evaluation of the ablation zone was also performed at 24 h and at 7 days after RF ablation. Results From ex vivo experiments, the ablation volume was found to increase significantly when RF energy was applied with the chilled water circulation. Results of in vivo experiments demonstrate that the ablation volume reached its maximum value when RF power of 7 W was applied (532.3 ± 110.3 mm 3 ). Histopathological examination showed delineated coagulation necrosis at 24 h after RF ablation, which clarified the ablation zone border. Fibrotic change was also observed at 7 days after RF ablation. Conclusion RF ablation using a 21-gauge electrode produced coagulation necrosis in the rat liver. The ablation volume became maximum when RF power of 7 W was applied with chilled water circulation.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0174-1551
1432-086X
DOI:10.1007/s00270-020-02660-w