NightBalance Sleep Position Treatment Device Versus Auto-Adjusting Positive Airway Pressure for Treatment of Positional Obstructive Sleep Apnea
Compare treatment efficacy and objective adherence between the NightBalance sleep position treatment (SPT) device and auto-adjusting positive airway pressure (APAP) in patients with exclusive positional obstructive sleep apnea (ePOSA) defined as a supine apnea-hypopnea index (sAHI) ≥ 2 times the non...
Saved in:
Published in | Journal of clinical sleep medicine Vol. 15; no. 7; pp. 947 - 956 |
---|---|
Main Authors | , , , , , , , |
Format | Journal Article |
Language | English |
Published |
United States
American Academy of Sleep Medicine
15.07.2019
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Compare treatment efficacy and objective adherence between the NightBalance sleep position treatment (SPT) device and auto-adjusting positive airway pressure (APAP) in patients with exclusive positional obstructive sleep apnea (ePOSA) defined as a supine apnea-hypopnea index (sAHI) ≥ 2 times the nonsupine AHI (nsAHI) and a nsAHI < 10 events/h.
This prospective multicenter randomized crossover trial enrolled treatment naive participants with ePOSA (AHI ≥ 15 events/h and nsAHI < 10 events/h) or (AHI > 10 and < 15 events/h with daytime sleepiness and nsAH < 5 events/h). Polysomnography and objective adherence determination (device data) were performed at the end of each 6-week treatment. Patient device preference was determined at the end of the study.
A total of 117 participants were randomized (58 SPT first, 59 APAP first). Of these, 112 started treatment with the second device (adherence cohort) and 110 completed the study (AHI cohort). The AHI on SPT was higher (mean ± standard deviation, 7.29 ± 6.8 versus 3.71 ± 5.1 events/h,
< .001). The mean AHI difference (SPT-APAP) was 3.58 events/h with a one sided 95% confidence interval upper bound of 4.96 events/h (< the prestudy noninferiority margin of 5 events/h). The average nightly adherence (all nights) was greater on SPT (345.3 ± 111.22 versus 286.98 ± 128.9 minutes,
< .0001). Participants found the SPT to be more comfortable and easier to use and 53% reported a preference for SPT assuming both devices were equally effective.
Treatment with SPT resulted in non-inferior treatment efficacy and greater adherence compared to APAP in ePOSA suggesting that SPT is an effective treatment for this group.
Registry: ClinicalTrials.gov; Title: The POSAtive Study: Study for the Treatment of Positional Obstructive Sleep Apnea; Identifier: NCT03061071; URL: https://clinicaltrials.gov/ct2/show/NCT03061071. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-News-2 ObjectType-Feature-3 content type line 23 |
ISSN: | 1550-9389 1550-9397 |
DOI: | 10.5664/jcsm.7868 |