Fractional moving blood volume: estimation with power Doppler US
To estimate the fraction of moving blood in tissue with power Doppler ultrasound (US). Power Doppler US measurements of moving scatterers in a flow tube were made as a function of successive dilutions of the perfusate. Measurements were normalized relative to the maximum Doppler power in the center...
Saved in:
Published in | Radiology Vol. 197; no. 1; p. 183 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | English |
Published |
United States
01.10.1995
|
Subjects | |
Online Access | Get more information |
Cover
Loading…
Summary: | To estimate the fraction of moving blood in tissue with power Doppler ultrasound (US).
Power Doppler US measurements of moving scatterers in a flow tube were made as a function of successive dilutions of the perfusate. Measurements were normalized relative to the maximum Doppler power in the center of the flow tube at the highest concentration and were used to calculate the fractional dilution of the perfusate for each run with each dilution used to represent increasing amounts of non-moving soft tissue in the sample volume. The technique was also applied to two clinical examples.
Successive dilutions of the perfusate in the flow experiment showed a monotonic, linear decrease in the Doppler power as a function of dilution.
The power Doppler US technique has the potential to more accurately estimate alterations in blood flow and has the advantage of being a continuous parameter that can be depth normalized. |
---|---|