Stray light correction of array spectroradiometer measurement in ultraviolet

For most of the array spectroradiometer, stray light is significant in UV band. Stray light correction of a UV array spectroradiometer is investigated using optical filters. If a group of filters with continuous bandpass are chosen, stray light contribution due to all the bands can be obtained using...

Full description

Saved in:
Bibliographic Details
Published inJournal of physics. Conference series Vol. 972; no. 1; pp. 12022 - 12027
Main Authors Wu, Zhifeng, Dai, Caihong, Wang, Yanfei, Li, Ling
Format Journal Article
LanguageEnglish
Published Bristol IOP Publishing 01.02.2018
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:For most of the array spectroradiometer, stray light is significant in UV band. Stray light correction of a UV array spectroradiometer is investigated using optical filters. If a group of filters with continuous bandpass are chosen, stray light contribution due to all the bands can be obtained using a numerical algorithm. The array spectroradiometer with the stray light corrected is used to measure the spectral irradiance of several UV lamps. The measurement results are compared to a double monochromator spectroradiometer. When xenon lamp is the array spectroradiometer calibration lamp, after stray light correction, the difference can be improved from nearly 10% to 2.0% in UVC band. When tungsten lamp is the calibration lamp, the difference can be improved from around 90% to less than 20%.
ISSN:1742-6588
1742-6596
DOI:10.1088/1742-6596/972/1/012022