An electrooptic probe to determine internal electric fields in a piezoelectric transformer
A technique using the electrooptic effect to determine the output voltage of an optically clear LiNbO(3) piezoelectric transformer was developed and explored. A brief mathematical description of the solution is provided, as well as experimental data demonstrating a linear response under ac resonant...
Saved in:
Published in | Review of scientific instruments Vol. 83; no. 2; p. 025106 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
United States
01.02.2012
|
Online Access | Get more information |
Cover
Loading…
Summary: | A technique using the electrooptic effect to determine the output voltage of an optically clear LiNbO(3) piezoelectric transformer was developed and explored. A brief mathematical description of the solution is provided, as well as experimental data demonstrating a linear response under ac resonant operating conditions. A technique to calibrate the diagnostic was developed and is described. Finally, a sensitivity analysis of the electrooptic response to variations in angular alignment between the LiNbO(3) transformer and the laser probe are discussed. |
---|---|
ISSN: | 1089-7623 |
DOI: | 10.1063/1.3682783 |