Characterization of device-related interruptions in minimally invasive surgery: need for intraoperative data and effective mitigation strategies

Background The burden of device-related interruptions is expected to increase as modern surgical practices adopt complex minimally invasive surgery devices. Currently, there is a paucity of empiric data that examined the nature of device-related interruptions using comprehensive intraoperative data....

Full description

Saved in:
Bibliographic Details
Published inSurgical endoscopy Vol. 33; no. 3; pp. 717 - 723
Main Authors Jung, James J., Kashfi, Arash, Sharma, Sahil, Grantcharov, Teodor
Format Journal Article
LanguageEnglish
Published New York Springer US 01.03.2019
Springer Nature B.V
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Background The burden of device-related interruptions is expected to increase as modern surgical practices adopt complex minimally invasive surgery devices. Currently, there is a paucity of empiric data that examined the nature of device-related interruptions using comprehensive intraoperative data. Methods We performed a cross-sectional study of consecutive elective laparoscopic general surgery cases performed in one operating room (OR) at a referral center between April 2014 and April 2016. The included cases were directly observed using a comprehensive multiport data recorder called the OR Black Box. The data were synchronized, encrypted, and reviewed by expert surgeon assessors. The assessors characterized device-related interruptions that occurred during operations. The prevalence of the cases with device-related interruptions was calculated. Device-related interruptions were classified into a priori categories of (1) absent/wrong device; (2) improper assembly; (3) loss of sterility; (4) disconnection; and (5) device failure. Results In a cohort of 210 cases, 64 (30%) had at least one device-related interruption. Sleeve gastrectomy (52%) and oncologic gastrectomy (43%) procedures experienced the highest prevalence of device-related interruptions. Device failure was the most frequently chosen category with laparoscopic staplers implicated in more than half of these failures. Three failure modes were described for laparoscopic stapler, of which stapler malfunction (46%) was the most common. Conclusions Device-related interruptions occurred frequently in the OR and could be characterized into one of the five categories. Understanding the nature of the device-related interruptions can help guide implementation of safety interventions and user training in the future.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0930-2794
1432-2218
DOI:10.1007/s00464-018-6254-5