Multicenter analytical performance evaluation of the Elecsys® proBNP assay

The purpose of this multicenter study was to evaluate the technical performance of the automated Elecsys proBNP (brain natriuretic peptide) assay, which is indicated as an aid in the diagnosis of individuals suspected of having congestive heart failure. The Elecsys proBNP assay is an electrochemilum...

Full description

Saved in:
Bibliographic Details
Published inClinical chemistry and laboratory medicine Vol. 42; no. 8; pp. 965 - 972
Main Authors Sokoll, Lori J., Baum, Hannsjörg, Collinson, Paul O., Gurr, Eberhard, Haass, Markus, Luthe, Hilmar, Morton, James J., Nowatzke, William, Zingler, Christiana
Format Journal Article
LanguageEnglish
Published Berlin De Gruyter 01.01.2004
New York, NY Walter de Gruyter
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The purpose of this multicenter study was to evaluate the technical performance of the automated Elecsys proBNP (brain natriuretic peptide) assay, which is indicated as an aid in the diagnosis of individuals suspected of having congestive heart failure. The Elecsys proBNP assay is an electrochemiluminescent immunoassay employing two polyclonal NT-proBNP-specific antibodies in a sandwich test format. The study was performed on the three Elecsys analyzers (E 1010, E 2010, and E 170) at eight different sites world-wide. Within- and total precision were ≤3%, with total precision slightly higher on the Elecsys E 170 instrument with multiple modules. Reproducibility among sites and platforms was <5%. Precision at particularly low NT-proBNP concentrations was assessed down to approximately 25 pg/ml with CVs of 12.6% at 29.2 pg/ml and 9.6% at 38.5 pg/ml for the Elecsys 1010/2010 and E 170, respectively. Linearity was evaluated up to 25,000 pg/ml with a sample-based non-linear response observed with recoveries of <90% for proBNP concentrations <10 000 pg/ml. Slopes ranged between 0.92 and 1.02 and intercepts from –5.3 to 10.4 pg/ml (r≥0.998) among the three types of analyzers. Slopes were 4.95 and 4.53 in comparison to the Biosite Triage and Shionogi BNP assays. There was no assay interference, and no effect of barrier gels, tube composition, or freeze-thaw. NT-proBNP concentrations in EDTA plasma were up to 10% lower than in serum or heparinized plasma and the analyte was stable at 4°C for up to 72 hours (the maximum time tested). There was no circadian rhythm in normal subjects or congestive heart failure patients and there was no effect of drawing position. In summary, the Elecsys proBNP assay exhibits good technical performance and is suitable for use in routine clinical laboratories to aid in the diagnosis of congestive heart failure.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1434-6621
1437-4331
DOI:10.1515/CCLM.2004.157