Continuous measurements by impedance of haematocrit and plasma volume variations during dialysis
A technique for continuous measurements of haematocrit and plasma volume in the arterial line of dialysed patients has been tested in vitro and in vivo. This method uses impedance measurements at 5 kHz and requires a single haematocrit measurement. It relies on two assumptions: that plasma resistivi...
Saved in:
Published in | Medical & biological engineering & computing Vol. 35; no. 3; pp. 167 - 171 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
Heidelberg
Springer
01.05.1997
Springer Nature B.V |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | A technique for continuous measurements of haematocrit and plasma volume in the arterial line of dialysed patients has been tested in vitro and in vivo. This method uses impedance measurements at 5 kHz and requires a single haematocrit measurement. It relies on two assumptions: that plasma resistivity does not change during dialysis and that blood resistivity obeys Hanai's model. Both assumptions are verified during in vitro tests. Haematocrits measured in vivo by this method are found to be in good agreement with direct measurements from blood samples. The haematocrit variation is then used to monitor changes in plasma volume, assuming conservation of erythrocyte volume. In addition, it is possible to obtain the variation in interstitial volume by combining these data with body impedance measurements. |
---|---|