Reel syndrome—not a twiddler variant
We report three cases of an unusual form of “reel syndrome” characterized by isolated, reeling dislodgement of a single lead in patients with dual-chamber or biventricular devices. One of these patients presented with worsening heart failure due to loss of left ventricular pacing and the others were...
Saved in:
Published in | Journal of interventional cardiac electrophysiology Vol. 23; no. 3; pp. 243 - 246 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
Boston
Springer US
01.12.2008
Springer Nature B.V |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | We report three cases of an unusual form of “reel syndrome” characterized by isolated, reeling dislodgement of a single lead in patients with dual-chamber or biventricular devices. One of these patients presented with worsening heart failure due to loss of left ventricular pacing and the others were detected incidentally during scheduled device checks. We suspect that a ratchet mechanism was probably responsible for this and that this type of dislodgement is not a twiddler variant. We propose a simple solution for prevention. |
---|---|
Bibliography: | ObjectType-Case Study-2 SourceType-Scholarly Journals-1 ObjectType-Feature-4 content type line 23 ObjectType-Report-1 ObjectType-Article-3 |
ISSN: | 1383-875X 1572-8595 |
DOI: | 10.1007/s10840-008-9288-0 |