Performance of LED fluorescence microscopy for the detection of tuberculosis in Rwanda using Zeiss Primo Star
INTRODUCTION: Ziehl-Neelsen (ZN) bright-field microscopy is time-consuming, with poor sensitivity, even under optimal conditions. Introduction of Primo StariLED fluorescent microscopy (FM) may improve TB case finding at referral hospitals in Rwanda.The study aimed to determine the acceptability and...
Saved in:
Published in | The Pan African medical journal Vol. 21; no. 198 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
The Pan African Medical Journal
14.07.2015
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | INTRODUCTION: Ziehl-Neelsen (ZN) bright-field microscopy is time-consuming, with poor sensitivity, even under optimal conditions. Introduction of Primo StariLED fluorescent microscopy (FM) may improve TB case finding at referral hospitals in Rwanda.The study aimed to determine the acceptability and effectiveness of iLED in a low resource setting. METHODS: between June 2009 and May 2010, the Rwandan TBProgramand National Reference Laboratorycarried out demonstration studies with iLED at a referral hospital in the capital,Kigali, and a rural district hospital in Nyamata, takingconventional FM as Gold Standard. RESULTS: agreement between the iLED and rechecking at the Reference Laboratory were deemed almost perfect (kappa = 0.81-1.00) across three of four site-phase combinations. The exception was Nyamata District Hospital during the validation phase, which was deemed substantial agreement (kappa = 0.61-0.80). However, the 100% concordance at both demonstration sites during the continuation phase shows technicians' rapid command of the new iLED microscope in a relatively short time. The lower overall positivity rate obtained in the rural clinicis not related to the performance of the microscope (or technicians), but is attributable to a significant increase in total number of patients and samples screened through active case finding. CONCLUSIONS: laboratory technicians demonstrated high acceptance ofiLED. Additionally, fluorescent microscopy reduces the time necessary for examination by more than half. The high level of agreement between iLED and FMduring implementation in both sites provides initial evidence for iLED to replace current methods. |
---|---|
ISSN: | 1937-8688 1937-8688 |
DOI: | 10.11604/pamj.2015.21.198.5776 |