Performance of a novel BD Stem Cell Enumeration Kit on two flow cytometry systems
Summary Introduction Hematopoietic stem cell transplantation, which requires accurate enumeration of stem cells, is routinely used in clinical settings. Flow cytometry provides a qualitative and quantitative assessment of CD34+ cells. Precision, linearity, and stability of the novel BD™ Stem Cell En...
Saved in:
Published in | International journal of laboratory hematology Vol. 35; no. 4; pp. 393 - 399 |
---|---|
Main Authors | , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
England
Blackwell Publishing Ltd
01.08.2013
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Summary
Introduction
Hematopoietic stem cell transplantation, which requires accurate enumeration of stem cells, is routinely used in clinical settings. Flow cytometry provides a qualitative and quantitative assessment of CD34+ cells. Precision, linearity, and stability of the novel BD™ Stem Cell Enumeration (SCE) Kit were evaluated on two flow cytometry platforms using a modified ISHAGE gating strategy and including a viability dye for data acquisition and analysis.
Methods
Precision and linearity were evaluated on BD FACSCanto™ II and BD FACSCalibur™ systems. Stability was evaluated on the BD FACSCanto II system. Precision was tested using both high and low controls. Linearity was evaluated using dilutions from CD34+ cell pools, while stability was evaluated using fresh leukapheresis specimens.
Results
Both systems showed precision with limited variability in absolute counts and percentages of viable CD34+ cells. The linearity range of viable CD34+ cells in both systems was established at 0–1000 cells/μL, showing a linear relationship (R2 = 0.99). Stability of CD34+ cells in mobilized leukapheresis samples was confirmed up to 24 h after collection and up to 60 min after the end of stain/lyse procedures.
Conclusion
The BD SCE Kit on both flow cytometry systems shows consistent and reproducible results. |
---|---|