Inappropriate Defibrillator Shocks from Depolarization—Repolarization Mismatch in a Patient with Hypertrophy Cardiomyopathy

Despite wide use of dedicated bipolar sensing electrodes in implantable cardioverter‐defibrillator (ICD) systems, oversensing occasionally occurs, leading to unwarranted shocks or antitachycardia pacing. This case report highlights an individual with hypertrophic cardiomyopathy (HCM) who experienced...

Full description

Saved in:
Bibliographic Details
Published inPacing and clinical electrophysiology Vol. 30; no. 11; pp. 1408 - 1411
Main Authors KAPA, SURAJ, CURWIN, JAY H., COYNE, ROBERT F., WINTERS, STEPHEN L.
Format Journal Article
LanguageEnglish
Published Malden, USA Blackwell Publishing Inc 01.11.2007
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Despite wide use of dedicated bipolar sensing electrodes in implantable cardioverter‐defibrillator (ICD) systems, oversensing occasionally occurs, leading to unwarranted shocks or antitachycardia pacing. This case report highlights an individual with hypertrophic cardiomyopathy (HCM) who experienced inappropriate shocks from oversensing of repolarization electrograms (T‐waves). During the implantation procedure, no excessive T‐wave amplitudes were detected during sinus rhythm, ventricular pacing, or induced ventricular fibrillation. T‐wave oversensing leading to shocks only developed after maturation of the lead–tissue interface. An adequate safety margin for discrimination between ventricular electrograms and T‐waves could not be assured. Thus, insertion of a new dedicated pacing‐sensing electrode was required. The degree to which intracardiac repolarization signals may be heightened in patients with HCM has not been investigated systematically. However, a relative decrease in the ventricular electrogram amplitude without a concomitant decline of the intracardiac T‐wave amplitude appears to have led to the problem in this patient. Special caution in technique and device selection with a particular emphasis on T‐wave sensing may be prudent when ICDs are implanted in individuals with HCM. Additional programmable variables may also be beneficial in such cases.
Bibliography:ObjectType-Case Study-2
SourceType-Scholarly Journals-1
ObjectType-Feature-4
content type line 23
ObjectType-Report-1
ObjectType-Article-3
ISSN:0147-8389
1540-8159
DOI:10.1111/j.1540-8159.2007.00881.x