Single-acquisition method for simultaneous determination of extrinsic gamma-camera sensitivity and spatial resolution

A new method for measuring simultaneously both the extrinsic sensitivity and spatial resolution of a gamma-camera in a single planar acquisition was implemented. A dual-purpose phantom (SR phantom; sensitivity/resolution) was developed, tested and the results compared with other conventional methods...

Full description

Saved in:
Bibliographic Details
Published inApplied radiation and isotopes Vol. 66; no. 1; pp. 44 - 49
Main Authors Santos, J.A.M., Sarmento, S., Alves, P., Torres, M.C., Bastos, A.L., Ponte, F.
Format Journal Article
LanguageEnglish
Published England Elsevier Ltd 2008
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:A new method for measuring simultaneously both the extrinsic sensitivity and spatial resolution of a gamma-camera in a single planar acquisition was implemented. A dual-purpose phantom (SR phantom; sensitivity/resolution) was developed, tested and the results compared with other conventional methods used for separate determination of these two important image quality parameters. The SR phantom yielded reproducible and accurate results, allowing an immediate visual inspection of the spatial resolution as well as the quantitative determination of the contrast for six different spatial frequencies. It also proved to be useful in the estimation of the modulation transfer function ( MTF) of the image formation collimator/detector system at six different frequencies and can be used to estimate the spatial resolution as function of the direction relative to the digital matrix of the detector.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0969-8043
1872-9800