Implantable Cardioverter Defibrillator T Wave Oversensing Caused by Hyperglycemia
KRISHEN, A., et al.: Implantable Cardioverter Defibrillator T Wave Oversensing Caused by Hyperglycemia. This case report highlights a previously unreported cause of T wave oversensing in a patient with an ICD and recent onset diabetes. Consistent T wave oversensing was observed at elevated serum glu...
Saved in:
Published in | Pacing and clinical electrophysiology Vol. 24; no. 11; pp. 1701 - 1703 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
Oxford, UK
Blackwell Futura Publishing, Inc
01.11.2001
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | KRISHEN, A., et al.: Implantable Cardioverter Defibrillator T Wave Oversensing Caused by Hyperglycemia. This case report highlights a previously unreported cause of T wave oversensing in a patient with an ICD and recent onset diabetes. Consistent T wave oversensing was observed at elevated serum glucose levels and this finding was reproduced with a glucose challenge. No T wave oversensing was seen during treatment of hyperglycemia. Alterations in serum chemistry may account for intermittent T wave oversensing in patients with ICDs. |
---|---|
Bibliography: | ObjectType-Case Study-2 SourceType-Scholarly Journals-1 ObjectType-Feature-4 content type line 23 ObjectType-Report-1 ObjectType-Article-3 |
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1046/j.1460-9592.2001.01701.x |