Quality Evaluation of Image Recording Strategies for Limited Angle Tomography

In this paper we propose a new method for evaluating image recording strategies for limited angle tomography. In limited angle tomography exact three-dimensional (3-D) reconstruction is not achievable. With this method a metric for the reachable reconstruction quality by defined X-ray source traject...

Full description

Saved in:
Bibliographic Details
Published inTsinghua science and technology Vol. 15; no. 1; pp. 25 - 29
Main Authors Stopp, Fabian, Winne, Christian, Jank, Emanuel, Keeve, Erwin
Format Journal Article
LanguageEnglish
Chinese
Published Elsevier Ltd 01.02.2010
Fraunhofer Institute for Production Systems and Design Technology,10587 Berlin
Department of Maxillofacial Surgery and Clinical Navigation,Charité - Universit(a)tsmedizin Berlin,13353 Berlin,Germany%Fraunhofer Institute for Production Systems and Design Technology,10587 Berlin,Germany%Department of Maxillofacial Surgery and Clinical Navigation,Charité - Universit(a)tsmedizin Berlin,13353 Berlin,Germany
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:In this paper we propose a new method for evaluating image recording strategies for limited angle tomography. In limited angle tomography exact three-dimensional (3-D) reconstruction is not achievable. With this method a metric for the reachable reconstruction quality by defined X-ray source trajectories is calculated. The result of our method is independent of reconstruction algorithms. Our approach is based on the gradients of the scanned volume and their grade of determinability. Compared to simulated reconstruction accuracy with simultaneous algebraic reconstruction techniques, the method of evaluation shows the same dependencies on X-ray source trajectories. By using the proposed method different source trajectories for a limited angle range are comparable with respect to the reachable reconstruction quality.
Bibliography:3-D reconstruction quality
11-3745/N
intraoperative imaging
TB852.1
X-ray source trajectories
limited angle tomography
TP391.41
3-D reconstruction quality; intraoperative imaging; limited angle tomography; X-ray source trajectories
ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Feature-1
content type line 23
ISSN:1007-0214
1878-7606
1007-0214
DOI:10.1016/S1007-0214(10)70004-3