Application of actuator-driven pulsed water jet in aneurysmal subarachnoid hemorrhage surgery: its effectiveness for dissection around ruptured aneurysmal walls and subarachnoid clot removal
In clipping surgery for aneurysmal subarachnoid hemorrhage (aSAH), critical steps include clot removal and dissection of aneurysms without premature rupture or brain injuries. To pursue this goal, a piezo actuator-driven pulsed water jet (ADPJ) system was introduced in this study. This study include...
Saved in:
Published in | Neurosurgical review Vol. 40; no. 3; pp. 485 - 493 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
Berlin/Heidelberg
Springer Berlin Heidelberg
01.07.2017
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | In clipping surgery for aneurysmal subarachnoid hemorrhage (aSAH), critical steps include clot removal and dissection of aneurysms without premature rupture or brain injuries. To pursue this goal, a piezo actuator-driven pulsed water jet (ADPJ) system was introduced in this study. This study included 42 patients, who suffered aSAH and underwent clipping surgery. Eleven patients underwent surgery with the assistance of the ADPJ system (ADPJ group). In the other 31 patients, surgery was performed without the ADPJ system (Control group). The ADPJ system was used for clot removal and aneurysmal dissection. The clinical impact of the ADPJ system was judged by comparing the rate of premature rupture, degree of clot removal, and clinical outcomes. Intraoperatively, a premature rupture was encountered in 18.2 and 25.8% of cases in the ADPJ and control groups, respectively. Although the differences were not statistically significant, intraoperative observation suggested that the ADPJ system was effective in clot removal and dissection of aneurysms in a safe manner. Computed tomography scans indicated the achievement of higher degrees of clot removal, especially when the ADPJ system was used for cases with preoperative clot volumes of more than 25 ml (
p
= 0.047, Mann–Whitney U test). Clinical outcomes, including incidence of postoperative brain injury or symptomatic vasospasm, were similar in both groups. We described our preliminary surgical results using the ADPJ system for aSAH. Although further study is needed, the ADPJ system was considered a safe and effective tool for clot removal and dissection of aneurysms. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0344-5607 1437-2320 |