Calibration of tracking systems in a surgical environment

The purpose of this paper was to assess to what extent an optical tracking system (OTS) used for position determination in computer-aided surgery (CAS) can be enhanced by combining it with a direct current (DC) driven electromagnetic tracking system (EMTS). The main advantage of the EMTS is the fact...

Full description

Saved in:
Bibliographic Details
Published inIEEE transactions on medical imaging Vol. 17; no. 5; pp. 737 - 742
Main Authors Birkfellner, W., Watzinger, F., Wanschitz, F., Ewers, R., Bergmann, H.
Format Journal Article
LanguageEnglish
Published United States IEEE 01.10.1998
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The purpose of this paper was to assess to what extent an optical tracking system (OTS) used for position determination in computer-aided surgery (CAS) can be enhanced by combining it with a direct current (DC) driven electromagnetic tracking system (EMTS). The main advantage of the EMTS is the fact that it is not dependent on a free line-of-sight. Unfortunately, the accuracy of the EMTS is highly affected by nearby ferromagnetic materials. The authors have explored to what extent the influence of the metallic equipment in the operating room (OR) can be compensated by collecting precise information on the nonlinear local error in the EMTS by using the OTS for setting up a calibration look-up table. After calibration of the EMTS and registration of the sensor systems in the OR the authors have found the average euclidean deviation in position readings between the DC tracker and the OTS reduced from 2.9/spl plusmn/1.0 mm to 2.1/spl plusmn/0.8 mm within a half-sphere of 530-mm radius around the magnetic field emitter. Furthermore the authors have found the calibration to be stable after re-registration of the sensors under varying conditions such as different heights of the OR table and varying positions of the OR equipment over a longer time interval. These results encourage the further development of a hybrid magnetooptical tracker for computer-aided surgery where the electromagnetic tracker acts as an auxiliary source of position information for the optical system. Strategies for enhancing the reliability of the proposed hybrid magnetooptic tracker by detecting artifacts induced by mobile ferromagnetic objects such as surgical tools are discussed.
Bibliography:ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Feature-1
content type line 23
ObjectType-Article-1
ObjectType-Feature-2
ISSN:0278-0062
1558-254X
DOI:10.1109/42.736028