Evaluation of the Sysmex UF-1000i® urine flow cytometer in the diagnostic work-up of suspected urinary tract infection in a Dutch general hospital

Background: Automation and standardization of sediment analysis of urine samples by flow cytometry might serve as an alternative to labor-intensive laboratory methods, such as microscopic examination and culture. The Sysmex UF-1000i is a urine flow cytometer that uses two separate channels for count...

Full description

Saved in:
Bibliographic Details
Published inClinical chemistry and laboratory medicine Vol. 48; no. 12; pp. 1765 - 1771
Main Authors van der Zwet, Wil C., Hessels, Jan, Canbolat, Fatih, Deckers, Martine M.L.
Format Journal Article
LanguageEnglish
Published Berlin Walter de Gruyter 01.12.2010
De Gruyter
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Background: Automation and standardization of sediment analysis of urine samples by flow cytometry might serve as an alternative to labor-intensive laboratory methods, such as microscopic examination and culture. The Sysmex UF-1000i is a urine flow cytometer that uses two separate channels for counting blood cells and bacteria. Methods: In this study, 358 urine samples were analyzed with the Sysmex UF-1000i in parallel with manual microscopy, Gram stain and bacterial culture, the latter considered the gold standard. Results: Reproducibility for detection of white and red blood cells and bacteria was good, while detection of yeast proved unreliable. Depending on the definition of urinary tract in-fection (UTI) used, the negative predictive value and the per-centage of false-negative results were 100% and 0% [UTI ≥ 105 colony forming units (CFU)/mL] and 99% and 1.3%, (UTI ≥104 CFU/mL), respectively. Pre-screening with the Sysmex UF-1000i would have resulted in a reduction of bacterial culture by 42%. Carry over of bacteria between consecutive samples due to the use of fixed sample needle was observed, but did not result in false-positive interpretation of Sysmex UF-1000i results. Because of the occurrence of carry over, samples that have been analyzed with the Sysmex UF-1000i cannot be used for subsequent urine culture. Conclusions: In conclusion, the Sysmex UF-1000i offers the possibility for screening high numbers of urine samples in a fast and standardized way, resulting in a reduction in workload and speeding the diagnostic process. It is not recommended for use in complicated patient populations, such as neutropenic patients and patients in whom yeast infection is suspected. Clin Chem Lab Med 2010;48:1765–71.
Bibliography:istex:45DCD2885EEA77FD65FB056056873A13F0F17B94
ark:/67375/QT4-BMJ7P55F-W
cclm.2010.342.pdf
ArticleID:cclm.2010.342
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1434-6621
1437-4331
DOI:10.1515/CCLM.2010.342