Comparison of CK-MB fractionation by antibody inhibition reaction and DEAE-Sephadex chromatography
1. CK-MB isoenzyme fractionation has been compared between DEAE-sephadex chromatography and the CK-M-subunit inhibiting antibody procedure. 2. It was found that the inhibiting antibody method generally gave lower values than the chromatography procedure. 3. The above difference can be partially attr...
Saved in:
Published in | Clinical biochemistry Vol. 11; no. 4; pp. 175 - 178 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
United States
Elsevier Inc
01.01.1978
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | 1.
CK-MB isoenzyme fractionation has been compared between DEAE-sephadex chromatography and the CK-M-subunit inhibiting antibody procedure.
2.
It was found that the inhibiting antibody method generally gave lower values than the chromatography procedure.
3.
The above difference can be partially attributed to inhibitory factors in the antibody system.
4.
It was concluded that although there was general agreement between the two procedures, the CK-M-subunit inhibiting antibody procedure should be viewed with caution at low creatine kinase activities, and should never be done without the appropriate reagent blanks. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0009-9120 1873-2933 |
DOI: | 10.1016/S0009-9120(78)90451-4 |