Clinical safety and performance of a MRI conditional pacing system in patients undergoing cardiac MRI
Background Utilization of cardiac magnetic resonance imaging (cMRI) as an imaging modality in clinical practice is rapidly increasing. More evidence from randomized studies establishing clinical safety and performance of pacing systems in patients undergoing a cMRI scan is needed. Objectives The pur...
Saved in:
Published in | Pacing and clinical electrophysiology Vol. 40; no. 12; pp. 1389 - 1395 |
---|---|
Main Authors | , , , , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
United States
Wiley Subscription Services, Inc
01.12.2017
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Background
Utilization of cardiac magnetic resonance imaging (cMRI) as an imaging modality in clinical practice is rapidly increasing. More evidence from randomized studies establishing clinical safety and performance of pacing systems in patients undergoing a cMRI scan is needed.
Objectives
The purpose of this prospective, multicenter, randomized study was to demonstrate safety and efficacy of the Accent MRI™ conditional pacing systems (St. Jude Medical, St. Paul, MN, USA) in patients undergoing cMRI scan.
Methods
Patients (n = 283) indicated for dual‐chamber pacemaker implant were randomized to either the MRI Group (MG) (n = 140) or the Control Group (CG) (n = 143) after successful device implantation of the Accent MRI™ system. Clinical evaluation and device interrogation were performed at pre‐ and post‐MRI scan, and 1 month post‐MRI for all patients. At 9–12 weeks postimplant, patients in MG underwent a nondiagnostic cMRI scan at 1.5 Tesla (T), while patients in CG underwent device interrogation and clinical evaluation twice with a 45‐minute waiting period in between. The safety endpoint was freedom from MRI scan‐related complications and that for efficacy was significant changes in right atrial/ventricular capture threshold and sensing amplitude between right before MRI, immediately after MRI, and 1 month post‐MRI.
Results
Results showed 100% freedom from MRI scan‐related complications. There were no significant changes in device performance between pre‐MRI scan and 1 month post‐MRI scan time points in both study groups.
Conclusions
cMRI scanning with 1.5 T scanners is safe in patients implanted with the Accent MRI™ conditional pacing system and has no significant effect on the electrical parameters of the device and leads. |
---|---|
Bibliography: | Clinicaltrials.gov This trial received funding from Abbott, formerly St. Jude Medical. Funding Identifier: NCT02041702 ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-News-2 ObjectType-Feature-3 content type line 23 |
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1111/pace.13232 |