Reaction time‐monitored patient‐maintained propofol sedation: a pilot study in oral surgery patients

Summary Previous volunteer studies of an effect‐site controlled patient‐maintained sedation system using propofol have demonstrated a risk of oversedation. We have incorporated a reaction time monitor into the handset to add an individualised patient‐feedback mechanism. This pilot study assessed if...

Full description

Saved in:
Bibliographic Details
Published inAnaesthesia Vol. 68; no. 7; pp. 760 - 764
Main Authors O'Brien, C., Urquhart, C. S., Allam, S, Anderson, K. J., Leitch, J. A., Macpherson, A., Kenny, G. N. C.
Format Journal Article
LanguageEnglish
Published England Blackwell Publishing Ltd 01.07.2013
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Summary Previous volunteer studies of an effect‐site controlled patient‐maintained sedation system using propofol have demonstrated a risk of oversedation. We have incorporated a reaction time monitor into the handset to add an individualised patient‐feedback mechanism. This pilot study assessed if the reaction time‐feedback modification would prove safe and effective in 20 healthy patients receiving sedation while undergoing oral surgery. All patients successfully sedated themselves without reaching any unsafe endpoints. All 20 maintained verbal contact throughout. The mean (SD) lowest peripheral blood oxygen saturation was 98.0 (2.1)% breathing room air. No patient required supplementary oxygen. The mean (SD) maximum effect‐site propofol concentration reached was 1.6 (0.5) μg.ml−1. The present system was found to be safe and effective, allowing oral surgery treatment under conscious sedation, but preventing oversedation.
Bibliography:http://www.anaesthesiacorrespondence.com
You can respond to this article at
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0003-2409
1365-2044