Design of the Double Ultrasonic Transducer for Nonlinearity Measurement of Acoustic Waves
In this paper, we propose and design a new ultrasonic probe that can effectively measure both the fundamental and the second-harmonic waves simultaneously. The transducer consists of two bonded Pb(Zr, Ti)O 3 (PZT) disks with the same properties and the same polarization, and a variable capacitor is...
Saved in:
Published in | Japanese Journal of Applied Physics Vol. 40; no. 5S; p. 3652 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
2001
|
Online Access | Get full text |
Cover
Loading…
Summary: | In this paper, we propose and design a new ultrasonic probe that can effectively measure both the fundamental and the second-harmonic waves simultaneously. The transducer consists of two bonded Pb(Zr, Ti)O
3
(PZT) disks with the same properties and the same polarization, and a variable capacitor is connected to the electric terminals of one of the PZT disks for the simulation and experiment. It is shown that the second harmonic as well as the fundamental component can be efficiently received by controlling the capacitance. The result of the nonlinearity parameter B/A for ethanol measured using the double ultrasonic transducer shows good agreement with the reference value, and the B/A value for soybean is obtained. |
---|---|
ISSN: | 0021-4922 1347-4065 |