Adverse Local Tissue Reaction due to Mechanically Assisted Crevice Corrosion Presenting as Late Instability Following Metal-on-Polyethylene Total Hip Arthroplasty

Mechanically assisted crevice corrosion (MACC) at modular junctions can cause a spectrum of adverse local tissue reactions (ALTRs) in patients who have undergone total hip arthroplasty (THA). The purpose of this study is to describe the presentation, treatments, and related complications of a cohort...

Full description

Saved in:
Bibliographic Details
Published inThe Journal of arthroplasty Vol. 35; no. 9; pp. 2666 - 2670
Main Authors Hannon, Charles P., Cotter, Eric J., Cooper, H. John, Deirmengian, Carl A., Rodriguez, Jose A., Urban, Robert M., Paprosky, Wayne G., Jacobs, Joshua J.
Format Journal Article
LanguageEnglish
Published United States Elsevier Inc 01.09.2020
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Mechanically assisted crevice corrosion (MACC) at modular junctions can cause a spectrum of adverse local tissue reactions (ALTRs) in patients who have undergone total hip arthroplasty (THA). The purpose of this study is to describe the presentation, treatments, and related complications of a cohort of patients presenting with late instability following metal-on-polyethylene THA due to underlying MACC and ALTR. This multicenter retrospective case series presents 17 patients (12 women, mean age 62.6, range 42-73) presenting with late instability secondary to ALTR and MACC. All patients had a metal (Cobalt Chrome)-on-polyethylene bearing surface. Patients experienced a mean 2.7 dislocations (range 1-6) at mean 4.3 years (range 0.4-17.0) following their index surgery. Serum metal levels (n = 12) demonstrated a greater elevation of cobalt (mean 6.9, range 0.13-20.88 ng/mL) than chromium (mean 1.9, range 0.13-3.23 ng/mL). Patients were revised for instability at a mean of 6.8 years (range 2.1-19.4) following their index surgery. ALTR was encountered in every case and the modular head-neck junction demonstrated visible corrosion. An exchange of the CoCr head to a ceramic head with a titanium sleeve and placement of a constrained liner was performed for a majority of patients (n = 15, 88.2%). Five patients (29.4%) had complications postoperatively including peroneal palsy (n = 2), periprosthetic joint infection (n = 2), and ALTR recurrence (n = 1). Recurrent instability in the setting of otherwise well-positioned THA components and without another obvious cause should raise concern for ALTR as a potential underlying etiology.
ISSN:0883-5403
1532-8406
DOI:10.1016/j.arth.2020.04.024