Current 25-hydroxyvitamin D assays: Do they pass the test?
Vitamin D testing is becoming increasingly important with recent research demonstrating a correlation between vitamin D insufficiency and metabolic diseases, immunodeficiencies and other diseases. However, existing 25-hydroxyvitamin D (25OHD) assays lack comparability to the candidate reference meth...
Saved in:
Published in | Clinica chimica acta Vol. 413; no. 13-14; pp. 1127 - 1134 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
Netherlands
Elsevier B.V
11.07.2012
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Vitamin D testing is becoming increasingly important with recent research demonstrating a correlation between vitamin D insufficiency and metabolic diseases, immunodeficiencies and other diseases. However, existing 25-hydroxyvitamin D (25OHD) assays lack comparability to the candidate reference method, causing difficulties in diagnosis and monitoring of vitamin D deficiency.
We looked at the accuracy of 3 automated assays (Roche Diagnostics Elecsys® Total 25OHD assay, Abbott Architect® Total vitamin D assay, Advia Centaur® vitamin D Total assay) and Diasorin® Radioimmunoassay (RIA) compared to a routine laboratory Liquid Chromatography–Tandem Mass Spectrometry (LC–MS/MS).
The correlation based on Passing Bablok regression was good with the slopes between 0.95 and 1.31 and the intercepts between −3.24 and 3.68. However, a significant positive bias was observed using the Abbott Architect and the Diasorin RIA. Using published analytical goals of coefficient of variation (CV) <10% and bias <5%, most methods did not meet these criteria. Using measurement of uncertainty of 9%, most methods were able to meet criteria using quality control materials but not patient samples.
Inadequacies of these assay performances are contributed by differences in method of extraction of vitamin D from vitamin D binding protein, cross-reactivities to 25OHD2, 25OHD3 and other vitamin D metabolites, matrix interferences and a lack of standardization.
► Five 25-hydroxyvitamin D (25OHD) assays were compared. ► Four 25OHD assays were evaluated using accuracy and precision data. ► The imprecision of quality control samples for the immunoassays was all below 10%. ► The imprecision of pooled serum samples for immunoassays was between 2.7% and 22.8%. ► Based on stringent analytical goals, most 25OHD assays did not meet the standards. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0009-8981 1873-3492 |