Multi-Channel Trans-Impedance Leadforming for Cardiopulmonary Monitoring: Algorithm Development and Feasibility Assessment Using In Vivo Animal Data
Objective : The objectives of this study were to develop a multi-channel trans-impedance leadforming method for beat-to-beat stroke volume (SV) and breath-by-breath tidal volume (TV) measurements and assess its feasibility on an existing in vivo animal dataset. Methods: A deterministic leadforming a...
Saved in:
Published in | IEEE transactions on biomedical engineering Vol. 69; no. 6; pp. 1964 - 1974 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
United States
IEEE
01.06.2022
The Institute of Electrical and Electronics Engineers, Inc. (IEEE) |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Objective : The objectives of this study were to develop a multi-channel trans-impedance leadforming method for beat-to-beat stroke volume (SV) and breath-by-breath tidal volume (TV) measurements and assess its feasibility on an existing in vivo animal dataset. Methods: A deterministic leadforming algorithm was developed to extract a cardiac volume signal (CVS) and a respiratory volume signal (RVS) from 208-channel trans-impedance data acquired every 20 ms by an electrical impedance tomography (EIT) device. SV EIT and TV EIT values were computed as a valley-to-peak value in the CVS and RVS, respectively. The method was applied to the existing dataset from five mechanically-ventilated pigs undergoing ten mini-fluid challenges. An invasive hemodynamic monitor was used in the arterial pressure-based cardiac output (APCO) mode to simultaneously measure SV APCO values while a mechanical ventilator provided TV Vent values. Results: The leadforming method could reliably extract the CVS and RVS from the 208-channel trans-impedance data measured with the EIT device, from which SV EIT and TV EIT were computed. The SV EIT and TV EIT values were comparable to those from the invasive hemodynamic monitor and mechanical ventilator. Using the data from 5 pigs and a simple calibration method to remove bias, the error in SV EIT and TV EIT was 9.5% and 5.4%, respectively. Conclusion: We developed a new leadforming method for the EIT device to robustly extract both SV and TV values in a deterministic fashion. Future animal and clinical studies are needed to validate this leadforming method in various subject populations. Significance: The leadforming method could be an integral component for a new cardiopulmonary monitor in the future to simultaneously measure SV and TV noninvasively, which would be beneficial to patients. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
ISSN: | 0018-9294 1558-2531 1558-2531 |
DOI: | 10.1109/TBME.2021.3132012 |